1 Indications And Usage


Ammonia N 13 Injection, USP is indicated for diagnostic Positron Emission Tomography (PET) imaging of the myocardium under rest or pharmacologic stress conditions to evaluate myocardial perfusion in patients with suspected or existing coronary artery disease.


Ammonia N 13 Injection, USP is a radioactive diagnostic agent for Positron Emission Tomography (PET) indicated for diagnostic PET imaging of the myocardium under rest or pharmacologic stress conditions to evaluate myocardial perfusion in patients with suspected or existing coronary artery disease (1).

2 Dosage And Administration



Rest Imaging Study (2.1):
  • •Aseptically withdraw Ammonia N 13 Injection from its container and administer 10 mCi—20 mCi (0.368 GBq—0.736 GBq) as a bolus through a catheter inserted into a large peripheral vein.
  • •Start imaging 3 minutes after the injection and acquire images for a total of 10-20 minute

Stress Imaging Study (2.2):
  • •If a rest imaging study is performed, begin the stress imaging study 40 minutes or more after the first Ammonia N 13 injection to allow for sufficient isotope decay.
  • •Administer a pharmacologic stress-inducing drug in accordance with its labeling.
  • •Aseptically withdraw Ammonia N 13 Injection from its container and administer 10 mCi—20 mCi (0.368 GBq—0.736 GBq) of Ammonia N 13 Injection as a bolus at 8 minutes  after the administration of the pharmacologic stress-inducing drug.
  • •Start imaging 3 minutes after the Ammonia N 13 Injection and acquire images for a total of 10-20 minutes.

Patient Preparation (2.3):
  • •To increase renal clearance of radioactivity and to minimize radiation dose to the bladder, hydrate the patient before the procedure and encourage voiding as soon as each image acquisition is completed and as often as possible thereafter for at least one hour.

2.1 Rest Imaging Study

  • •Aseptically withdraw Ammonia N 13 Injection from its container and administer 10 mCi-20 mCi (0.368 GBq-0.736 GBq) as a bolus through a catheter inserted into a large peripheral vein.
  • •Start imaging 3 minutes after the injection and acquire images for a total of 10-20 minutes.

2.2 Stress Imaging Study

  • •If a rest imaging study is performed, begin the stress imaging study 40 minutes or more after the first Ammonia N 13 injection to allow sufficient isotope decay.
  • •Administer a pharmacologic stress-inducing drug in accordance with its labeling.
  • •Aseptically withdraw Ammonia N 13 Injection from its container and administer 10 mCi—20 mCi (0.368 GBq—0.736 GBq) of Ammonia N 13 Injection as a bolus at 8 minutes after the administration of the pharmacologic stress-inducing drug.
  • •Start imaging 3 minutes after the Ammonia N 13 Injection and acquire images for a total of 10-20 minutes.

2.3 Patient Preparation


To increase renal clearance of radioactivity and to minimize radiation dose to the bladder, ensure that the patient is well hydrated before the procedure  and encourage voiding as soon as a study is completed and as often as possible thereafter for at least one hour.

2.4 Radiation Dosimetry


The converted radiation absorbed doses in rem/mCi are shown in Table 1. These estimates are calculated from the Task Group of Committee 2 of the International Commission on Radiation Protection.
Table 1. N13 Absorbed Radiation Dose Per Unit Activity (rem/mCi) for Adults and Pediatric Groups
Organ Age (Years)
Adult 15 10 5 1
* Upper large intestine, ** Lower large intestine

Adrenals

0.0085

0.0096

0.016

0.025

0.048

Bladder Wall

0.030

0.037

0.056

0.089

0.17

Bone Surfaces

0.0059

0.0070

0.011

0.019

0.037

Brain

0.016

0.016

0.017

0.019

0.027

Breast

0.0067

0.0067

0.010

0.017

0.033

Stomach Wall

0.0063

0.0078

0.012

0.019

0.037

Small Intestine

0.0067

0.0081

0.013

0.021

0.041

*ULI

0.0067

0.0078

0.013

0.021

0.037

**LLI

0.0070

0.0078

0.013

0.020

0.037

Heart

0.0078

0.0096

0.015

0.023

0.041

Kidneys

0.017

0.021

0.031

0.048

0.089

Liver

0.015

0.018

0.029

0.044

0.085

Lungs

0.0093

0.011

0.018

0.029

0.056

Ovaries

0.0063

0.0085

0.014

0.021

0.041

Pancreas

0.0070

0.0085

0.014

0.021

0.041

Red Marrow

0.0063

0.0078

0.012

0.020

0.037

Spleen

0.0093

0.011

0.019

0.030

0.056

Testes

0.0067

0.0070

0.011

0.018

0.035

Thyroid

0.0063

0.0081

0.013

0.021

0.041

Uterus

0.0070

0.0089

0.014

0.023

0.041

Other Tissues

0.0059

0.0070

0.011

0.018

0.035

2.5 Drug Handling

  • •Inspect Ammonia N 13 Injection visually for particulate matter and discoloration before administration, whenever solution and container permit.
  • •Do not administer Ammonia N 13 Injection containing particulate matter or discoloration; dispose of these unacceptable or unused preparations in a safe manner, in compliance with applicable regulations.
  • •Wear waterproof gloves and effective shielding when handling Ammonia N 13 Injection.
  • •Use aseptic technique to maintain sterility during all operation involved in the manipulation and administration of Ammonia N 13 Injection. The contents of each vial are sterile and non-pyrogenic.
  • •Use appropriate safety measures, including shielding, consistent with proper patient management to avoid unnecessary radiation exposure to the patient, occupational workers, clinical personnel, and other persons.
  • •Radiopharmaceuticals should be used by or under the control of physicians who are qualified by specific training and experience in the safe use and handling of radionuclides, and whose experience and training have been approved by the appropriate governmental agency authorized to license the use of radionuclides.
  • •Before administration of Ammonia N 13 Injection, assay the dose in a properly calibrated dose calibrator.

3 Dosage Forms And Strengths


Glass vial (30 mL) containing 0.138 GBq-1.387 GBq (3.75 mCi/mL-37.5 mCi/mL) of Ammonia N 13 Injection in aqueous 0.9% sodium chloride solution (approximately 10 mL volume) that is suitable for intravenous administration.


Glass vial containing 0.138 GBq-1.387 GBq (3.75 mCi/mL-37.5 mCi/mL) of Ammonia N 13 Injection in aqueous 0.9% sodium chloride solution (approximately 10mL volume) (3).

4 Contraindications


None.


None (4).

5 Warnings And Precautions



Ammonia N 13 Injection may increase the risk of cancer. Use the smallest dose necessary for imaging and ensure safe handling to protect the patient and health care worker (5.1).

5.1 Radiation Risks


Ammonia N 13 Injection may increase the risk of cancer. Use the smallest  dose necessary for imaging and ensure safe handling to protect the patient and health care worker [see Dosage and Administration (2.4)].

6 Adverse Reactions


No adverse reactions have been reported for Ammonia N 13 Injection based on a review of the published literature, publicly available reference sources, and adverse drug reaction reporting  systems. However, the completeness of these sources is not known.


No adverse reactions have been reported for Ammonia N 13 Injection based on a review of the published literature, publicly available reference sources, and adverse drug reaction reporting system (6).

To report SUSPECTED ADVERSE REACTIONS, contact Decatur Memorial Hospital at 800-826-2827 or FDA at 1-800-FDA-1088 or www.fda.gov/medwatch.

7 Drug Interactions


The possibility of interactions of Ammonia N 13 Injection with other drugs taken by patients undergoing PET imaging has not been studied.

8 Use In Specific Populations


  • •It is not known whether this drug is excreted in human milk. Alternatives to breastfeeding (e.g. using stored breast milk or infant formula) should be used for 2 hours (>10 half-lives of radioactive decay for N 13 isotope) after administration of Ammonia N 13 Injection (8.3).
  • •The safety and effectiveness of Ammonia N 13 Injection has been established in pediatric patients (8.4) 

8.1 Pregnancy


Pregnancy Category C

Animal reproduction studies have not been conducted with Ammonia N 13 Injection. It is also not known whether Ammonia N 13 Injection can cause fetal harm when administered to a pregnant woman or can affect reproduction capacity. Ammonia N 13 Injection should be given to a pregnant woman only if clearly needed.

8.3 Nursing Mothers


It is not known whether this drug is excreted in human milk. Because many drugs are excreted in human milk and because of the potential for radiation exposure to nursing infants from Ammonia N 13 Injection, use alternative infant nutrition sources (e.g. stored breast milk or infant formula) for 2 hours (>10 half-lives of radioactive decay for N 13 isotope) after administration of the drug or avoid use of the drug, taking into account the importance of the drug to the mother.

8.4 Pediatric Use


The safety and effectiveness of Ammonia N 13 Injection has been established in pediatric patients based on known metabolism of ammonia, radiation dosimetry in the pediatric population, and clinical studies in adults [see Dosage and Administration (2.4)].

11 Description


11.1 Chemical Characteristics


Ammonia N 13 Injection, USP is a positron emitting radiopharmaceutical that is used for diagnostic purposes in conjunction with positron emission tomography (PET) imaging. The active ingredient, [13 N] ammonia, has the molecular formula of 13NH3 with a molecular weight of 16.02, and has the following chemical structure:

Ammonia N 13 Injection is provided as a ready to use sterile, pyrogen-free, clear and colorless solution. Each mL of the solution contains between 0.138 GBq to 1.387 GBq (3.75 mCi to 37.5 mCi) of [13N] ammonia, at the end of synthesis (EOS) reference time, in 0.9% aqueous sodium chloride. The pH of the solution is between 4.5 to 7.5. The recommended dose of radioactivity (10-20 mCi) is associated with a theoretical mass dose of 0.5-1.0 picomoles of ammonia.

11.2 Physical Characteristics


Nitrogen N13 decays by emitting positron to Carbon C13 (stable) and has a physical half-life of 9.96 minutes. The principal photons useful for imaging are the dual 511 keV gamma photons that are produced and emitted simultaneously in opposite direction when the positron interacts with an electron (Table 2).
Table 2. Principal Radiation Emission Data for Nitrogen 13
* Produced by positron annihilation

Radiation/Emission

% Per Disintegration

Energy

Positron(β+)

100

1190 keV (Max.)

Gamma (±)*

200

511.0 keV

The specific gamma ray constant (point source air kerma coefficient) for nitrogen N 13 is 5.9 R/hr/mCi (1.39 x 10‑6 Gy/hr/kBq) at 1 cm. The half-value layer (HVL) of lead (Pb) for 511 keV photons is 4 mm.  Selected coefficients of attenuation are uled in Table 3 as a function of lead shield thickness. For example, the use of 39 mm thickness of lead will attenuate the external radiation by a factor of about 1000.
Table 3. Radiation Attenuation of 511keV Photons by lead (Pb) shielding

Shield thickness

(Pb) mm

Coefficient of

attenuation

4

0.50

8

0.25

13

0.10

26

0.01

39

0.001

52

0.0001

Table 4 uls fractions remaining at selected time intervals from the calibration time.  This information may be used to correct for physical decay  of the radionuclide.
Table 4. Physical Decay Chart for Nitrogen 13
* calibration time

Minutes

Fraction Remaining

0*

1.000

5

0.706

10

0.499

15

0.352

20

0.249

25

0.176

30

0.124

12 Clinical Pharmacology


12.1 Mechanism of Action


Ammonia N 13 Injection is a radio-labeled analog of ammonia that is distributed to all organs of the body after intravenous administration. It is extracted from the blood in the coronary capillaries into the myocardial cells where it is metabolized to glutamine N 13 and retained in the cells. The presence of ammonia N 13 and glutamine N 13 in the myocardium allows for PET imaging of the myocardium.

12.2 Pharmacodynamics


Following intravenous injection, ammonia N 13 enters the myocardium through the coronary arteries. The PET technique measures myocardial blood flow based on the assumption of a three compartmental disposition of intravenous ammonia N 13 in the myocardium. In this model, the value of the rate constant, which represents the delivery of blood to myocardium, and the fraction of ammonia N 13 extracted into the myocardial cells, is a measure of myocardial blood flow. Optima l PET imaging of the myocardium is generally achieved between 10 to 20 minutes after administration.

12.3 Pharmacokinetics


Following intravenous injection, Ammonia N 13 Injection is cleared from the blood with a biologic half-life of about 2.84 minutes (effective half-life of about 2.21 minutes). In the myocardium, its biologic half-life has been estimated to be less than 2 minutes (effective half-life less than 1.67 minutes).

The mass dose of Ammonia N 13 Injection is very small as compared to the normal range of ammonia in the blood (0.72-3.30 mg) in a healthy adult man [see Description (11.1)].

Plasma protein binding of ammonia N 13 or its N 13 metabolites has not been studied.

Ammonia N 13 undergoes a five-enzyme step metabolism in the liver to yield urea N 13 (the main circulating metabolite). It is also metabolized to glutamine N 13 (the main metabolite in tissues) by glutamine synthesis in the skeletal muscles, liver, brain, myocardium, and other organs. Other metabolites of ammonia N 13 include small amounts of N 13 amino acid anions (acidic amino acids) in the forms of glutamate N 13 or aspartate N 13.

Ammonia N 13 is eliminated from the body by urinary excretion mainly as urea N 13.

The pharmacokinetics of Ammonia N 13 Injection have not been studied in renally impaired, hepatically impaired, or pediatric patients.

13 Nonclinical Toxicology


13.1 Carcinogenesis, Mutagenesis, Impairment of Fertility


Long term animal studies have not been performed to evaluate the carcinogenic potential of Ammonia N 13 Injection. Genotoxicity assays and impairment of male and female fertility studies with Ammonia N 13 Injection have not been performed.

14 Clinical Studies


In a descriptive, prospective, blinded image interpretation study2 of adult patients with known or suspected coronary artery disease, myocardial perfusion deficits in stress and rest PET images obtained with Ammonia N 13 (N=111) or Rubidium 82 (N=82) were compared to changes in stenosis flow reserve (SFR) as determined by coronary angiography. The principal outcome of the study was the evaluation of PET defect severity relative to SFR.

PET perfusion defects at rest and stress for seven cardiac regions (anterior, apical, anteroseptal, posteroseptal, anterolateral, posterolateral, and inferior walls) were graded on a 0 to 5 scale defined as normal (0), possible (1), probable (2), mild (3), moderate (4), and severe (5) defects. Coronary angiograms were used to measure absolute and relative stenosis dimensions and to calculate stenosis flow reserve defined as the maximum value of flow at maximum coronary vasodilatation relative to rest flow under standardized hemodynamic conditions.

SFR scores ranged from 0 (total occlusion) to 5 (normal).

With increasing impairment of flow reserve, the subjective PET defect severity increased. A PET defect score of 2 or higher was positively correlated with flow reserve impairment (SFR<3).

15 References

  • 1.Annals of the ICRP. Publication 53. Radiation dose to patients from radiopharmaceuticals. New York: Pergamon Press, 1988.
  • 2.Demer, L.L.K.L.Gould, R.A.Goldstein, R.L.Kirkeeide, N.A.Mullani, R.W. Smalling, A.Nishikawa, and M.E.Merhige. Assessment of coronary artery disease severity by PET: Comparison with quantitative arteriography in 193 patients. Circulation 1989; 79: 825-3

16 How Supplied


Ammonia N 13 Injection, USP is packaged in 30 mL multiple dose glass vial containing between 1.11 GBq to 11.1 GBq (30 mCi to 300 mCi) of [13N] ammonia, at the end of synthesis (EOS) reference time, in 0.9% sodium chloride injection solution in approximately 10 mL volume. The recommended dose of radioactivity (10 mCi – 20 mCi) is associated with a theoretical mass dose of 0.5-1.0 picomoles of Ammonia.

NDC 73410-001-01

Storage

Store at 25°C (77°F); excursions permitted to 15-30°C (59-86°F). Use the solution within 60 minutes of the End of Synthesis (EOS) calibration.

17 Patient Counseling Information


17.1 Pre-Study Hydration


Instruct patients to drink plenty of water or other fluids (as tolerated) in the 4 hours before their PET study.

17.2 Post-Study Voiding


Instruct patients to void after completion of each image acquisition session and as often as possible for one hour after the PET scan ends.

17.3 Post-study Breastfeeding Avoidance


Instruct nursing patients to substitute stored breast milk or infant formula for breast milk for 2 hours after administration of Ammonia N 13 Injection.

09/2014


Manufactured by:

Decatur Memorial HospitalCenter for Advanced Molecular Medicine2300 North Edward Street, Suite 100Decatur, Illinois 62526

Distributed by: Decatur Memorial HospitalCenter for Advanced Molecular Medicine2300 North Edward Street, Suite 100Decatur, Illinois 62526

Principal Display Panel - Vial Label

Vial Label
Ammonia N 13Injection, USP

For Intravenous Use3.75 mCi/mL to37.5 mCi/mL @ EOS*

Lot #:

Rx Only

NDC# 73410-001-01

Sterile, Non-pyrogenic. Contains0.138 GBq to 1.387 GBq (3.75 mCi/mL to 37.5 mCi/mL) @ EOS* of no-carrier added Ammonia N13injection in 0.9% NaCl.

Multi-Dose Vial

Diagnostic - For IntravenousUse Only. Store @ 25°C (77°F);See Insert. Store upright in shielded container. Asepticallywithdraw/handle doses.

*EOS - End of Synthesis. Expires 60 minutes after EOS.Calculate correct dosage from date and time of calibration.Do not use if cloudy or contains particulate matter.(N13) Half-life = 9.96 minutes.

Manufactured by:

Decatur Memorial HospitalDecatur, IL 62526

Caution: Radioactive Material

DISCLAIMER:

"This tool does not provide medical advice, and is for informational and educational purposes only, and is not a substitute for professional medical advice, treatment or diagnosis. Call your doctor to receive medical advice. If you think you may have a medical emergency, please dial 911."

"Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service."

"This product uses publicly available data from the U.S. National Library of Medicine (NLM), National Institutes of Health, Department of Health and Human Services; NLM is not responsible for the product and does not endorse or recommend this or any other product."

PillSync may earn a commission via links on our site